The best b value of MR diffusion-weighted imaging for pancreatic cancer

GUO Li;YANG Da-kuan;YUAN Shu-guang;YAN Dong;WANG Jia-ping

Journal of China Clinic Medical Imaging ›› 2010, Vol. 21 ›› Issue (2) : 87-89.

Journal of China Clinic Medical Imaging ›› 2010, Vol. 21 ›› Issue (2) : 87-89.
论著

The best b value of MR diffusion-weighted imaging for pancreatic cancer

  • GUO Li, YANG Da-kuan, YUAN Shu-guang, YAN Dong, WANG Jia-ping
Author information +
History +

Abstract

Objective: To investigate the best b value for pancreatic cancer with DWI. Methods: Routine MR, DWI with different combination of b values [(0, 50), (0, 400), (0, 700), (0, 1000)s/mm2] were obtained in 73 cases(41 pancreatic cancer, 32 normal). The maps and the value of ADC in pancreatic cancer, tissue near the cancer, normal pancreatic tissue were analyzed and measured respectively. Results: The high image quality of ADC was the combination of (0, 400)s/mm2. There was a significant difference among tissue near the cancer, normal pancreatic tissue and pancreatic cancer. Conclusion: The (0,400)s/mm2 is the best b value for pancreatic cancer with DWI.

Key words

Pancreatic neoplasms / Diffusion magnetic resonance imaging

Cite this article

Download Citations
GUO Li;YANG Da-kuan;YUAN Shu-guang;YAN Dong;WANG Jia-ping. The best b value of MR diffusion-weighted imaging for pancreatic cancer[J]. Journal of China Clinic Medical Imaging. 2010, 21(2): 87-89

Accesses

Citation

Detail

Sections
Recommended

/