目的:评价梯度翻转脂肪抑制技术(LIPO)在3.0T磁共振胰腺弥散加权成像中的应用价值。方法:采用3.0T磁共振对20例正常志愿者分别采用LIPO技术和常规脂肪抑制技术行胰腺DWI扫描(b值800 s/mm2)。由两名经验丰富的影像科医师对所得DWI图像质量情况进行主观评分,并对胰腺表观扩散系数(ADC)值和信噪比(SNR)进行定量测量,应用SPSS 19.0统计学软件,采用单因素方差分析和LSD检验比较LIPO和压脂DWI序列图像中胰腺的ADC值、SNR及图像质量评分值。结果:正常胰腺在DWI LIPO和常规DWI序列中的ADC值无明显统计学差异(P=0.211);两者的SNR有明显统计学差异(P=0.000 2),且胰腺在DWI LIPO序列上的SNR明显高于常规DWI序列。与常规DWI序列比较,正常胰腺在DWI LIPO序列中伪影及图像模糊程度评分值均具有统计学差异(P值分别为0.001、0.009)。结论:LIPO技术可以在3.0T磁共振胰腺弥散加权成像扫描中提高图像质量。
Abstract
Objective: To evaluate the clinical value of gradient reversal fat suppression technology(LIPO) in diffusion-weighted MR imaging of normal pancreas at 3.0T. Methods: DWI-LIPO and conventional DWI sequence(b values 800 s/mm2) were performed among twenty volunteers on a 3.0T whole-body MRI system. The ADC value and image quality, including SNR, blurring and artifact level, was evaluated by two experienced radiologists. One-way analysis and LSD test were used to compare DWI-LIPO sequence with conventional DWI sequence through the software SPSS 19.0. Results: There was no statistical difference between two different DWI sequences(P=0.211) in ADC value of normal pancreas. Normal pancreas in DWI with LIPO has more higher SNR than conventional DWI(P=0.000 2). Compared with conventional DWI sequence, there was significantly statistical difference between the artifacts and image blurring degree scores in DWI with LIPO(P=0.001, 0.009). Conclusion: LIPO technology improves the image quality of pancreas DWI at 3.0T MR.
关键词
胰腺;磁共振成像 /
弥散
Key words
Pancreas /
Diffusion magnetic resonance imaging
{{custom_sec.title}}
{{custom_sec.title}}
{{custom_sec.content}}
参考文献
[1]刘豪,姚秀忠,陈财忠,等. 磁共振BLADE技术在肝脏成像中的临床应用[J]. 中国临床医学影像杂志,2014,25(6):406-408.
[2]姚秀忠,冯豪,陈文芳,等. 3.0T磁共振不同DWI序列在正常胰腺中的分析评价[J]. 中国临床医学影像杂志,2013,24(1):17-22.
[3]姚秀忠,许红莲,顾君英,等. 正常胰腺在3.0T MR不同DWI序列上的ADC值评价[J]. 放射学实践,2013,28(2):173-176.
[4]Nagy Z, Weiskopf N. Efficient fat suppression by slice-selection gradient reversal in twice-refocused diffusion encoding[J]. Magn Reson Med, 2008, 60(5): 1256-1260.
[5]Koh DW, Blackledge M, Burns S, et al. Combination of chemical suppression techniques for dual suppression of fat and silicone at diffusion-weighted MR imaging in women with breast implants[J]. Eur Radiol, 2012, 22(12): 2648-2653.
[6]Lee JS, Kim YK. Diffusion-weighted imaging of the liver at 3T using section-selection gradient reversal[J]. Clin Radiol, 2015, 70(4): 408-415.