自制水模在3.0T MRI颈部脂肪抑制序列扫描中的价值

汪 荣,朱大荣,韩志江,丁金旺

中国临床医学影像杂志 ›› 2017, Vol. 28 ›› Issue (3) : 166-168.

中国临床医学影像杂志 ›› 2017, Vol. 28 ›› Issue (3) : 166-168.
头颈部影像学

自制水模在3.0T MRI颈部脂肪抑制序列扫描中的价值

  • 汪  荣,朱大荣,韩志江,丁金旺
作者信息 +

The value of self-made water phantom for 3.0T MRI cervical fat-suppression sequence scan

  • WANG Rong, ZHU Da-rong, HAN Zhi-jiang, DING Jin-wang
Author information +
文章历史 +

摘要

目的:探讨自制水模在3.0T MRI颈部频率选择饱和法脂肪抑制(FS)序列扫描中的应用价值。方法:通过常规及水模序列,用相同参数对30例志愿者颈部进行FS快速自旋回波(TSE)T1WI和FS TSE T2WI扫描,对所得图像进行主观及客观评价,并比较分析。结果:主观评价法中,FS TSE T1WI分级在常规组和水模组分别为0级:1例和30例、1级:4例和0例、2级:8例和0例、3级:17例和0例,P均<0.05;FS TSE T2WI分级分别为0级:0例和1例、1级:1例和29例、2级:8例和0例、3级:21例和0例,P均<0.05。客观评价法中,FS TSE T1WI在下颌角、环状软骨、胸锁关节水平胸锁乳突肌信噪比中值在常规组和水模组分别为25.00(4.53,72.65)和103.72(99.77,108.04)、9.05(3.45,60.32)和118.80(106.59,123.94)、19.00(3.35,61.21)和101.74(87.38,110.29),P均<0.05;FS TSE T2WI像上分别为8.30(3.63,13.91)和15.26(12.68,17.50)、3.84(2.86,11.95)和15.06(12.76,16.88)、5.06(2.71,13.40)和14.46(11.47,15.68),P均<0.05。结论:在3.0T MRI颈部FS序列扫描中,应用水模可明显提高成像质量,并且操作简单,易于推广。

Abstract

Objective: To investigate the application value of self-made water phantom in 3.0T MRI cervical frequency selective saturated fat-suppression sequence scan. Methods: Necks of 30 volunteers were treated with FS TSE(rapid spin echo) T1WI and FS TSE T2WI scan by same parameters of routine and water phantom sequences, respectively. The obtained images were subjective and objectively evaluated for comparative analysis. Results: During subjective evaluation, FS TSE T1WI grading were 1 case and 30 cases at grade 0, 4 cases and 0 case at grade 1, 8 cases and 0 case at grade 2, 17 cases and 0 case at grade 3 in routine group and water phantom group, respectively(all P<0.05). FS TSE T2WI grading were 0 case and 1 case at grade 0, 1 case and 29 cases at grade 1, 8 cases and 0 case at grade 2, 21 cases and 0 case at grade 3 in routine group and water phantom group, respectively(all P<0.05). During objective evaluation, signal to noise ratio of FS TSE T1WI at mandibular angle, cricoid and sternocleidomastoid muscle at sternoclavicular joint level were 25.00(4.53, 72.65) and 103.72 (99.77, 108.04), 9.05(3.45, 60.32) and 118.80(106.59, 123.94), 19.00(3.35, 61.21) and 101.74(87.38, 110.29) in routine group and water phantom group, respectively(all P<0.05); signal to noise ratio of FS TSE T1WI were 8.30(3.63, 13.91) and 15.26(12.68, 17.50), 3.84(2.86, 11.95) and 15.06(12.76, 16.88), 5.06(2.71, 13.40) and 14.46(11.47, 15.68) in routine group and water phantom group, respectively(all P<0.05). Conclusion: For 3.0T MRI cervical FS sequence scan, water phantom can obviously improve image quality, be operated simply and easy to popularize.

关键词

/ 磁共振成像

Key words

Neck / Magnetic resonance imaging

引用本文

导出引用
汪 荣,朱大荣,韩志江,丁金旺. 自制水模在3.0T MRI颈部脂肪抑制序列扫描中的价值[J]. 中国临床医学影像杂志. 2017, 28(3): 166-168
WANG Rong, ZHU Da-rong, HAN Zhi-jiang, DING Jin-wang. The value of self-made water phantom for 3.0T MRI cervical fat-suppression sequence scan[J]. Journal of China Clinic Medical Imaging. 2017, 28(3): 166-168
中图分类号: R653    R445.2   

参考文献

[1]Lu H, Nagae-Poetscher LM, Golae X, et al. Routine clinical brain MRI sequences for use at 3.0 Tesla[J]. J Magn Reson Imaging, 2005, 22(1): 13-22.
[2]Alecci M, Collins CM, Smith MB, et al. Radio-frequency magnetic field mapping of a 3 Tesla birdcage coil: experimental and theoretical dependence on sample properties[J]. Magn Reson Med, 2001, 46(2): 379-385.
[3]郭勇,林伟,钱明珠,等. 臂丛神经磁共振IDEAL T2WI和CUBE Flex T2WI成像[J]. 中国医学影像学杂志,2011,19(7):499-502.
[4]肖建明,彭涛,陈志凡,等. 快速自旋回波与快速小角度激发抑脂序列在颈部增强扫描中的应用[J]. 中国医疗设备,2015,30(9):47-50.
[5]李颖,陈志安,乞文旭,等. GE 1.5T 430S MR频率饱和法与水脂分离法膝关节抑制脂肪图像比较分析[J]. 中国临床医学影像杂志,2013,24(12):873-876.
[6]吴婧,卢铃铨,顾建平,等. 磁共振脂肪抑制技术在骨与关节病变中的应用[J]. 医学影像学杂志,2013,23(1):146-150.
[7]Gold GE, Reeder SB, Yu H, et al. Articular cartilage of the knee: rapid three-dimensional MR imaging at 3.0T with IDEAL balanced steady-state free precession-initial experience[J]. Radiology, 2006, 240(2): 546-551.
[8]Karantanas AH, Zibis AH, Kitsoulis P. Fat-suppressed 3D-T1-weighted-echo planar imaging: comparison with fat-suppressed 3D-T1-weighted-gradient echo in imaging the cartilage of the knee[J]. Comput Med Imaging Graph, 2002, 26(3): 159-165.
[9]Stabler A, Spieker A, Bonel H, et al. Magnetic resonance imaging of the wrist-comparison of high resolution pulse sequences and different fat signal suppression techniques in cadavers[J]. Rofo, 2000, 172(2): 168-174.
[10]Bley TA, Wieben O, Francois CJ, et al. Fat and water magnetic resonance imaging[J]. J Magn Reson Imaging, 2010, 31(1): 4-18.
[11]Zhang Z, Song L, Meng Q, et al. Segmented echo planar MR imaging of the brachial plexus with inversion recovery magnetization preparation at 3.OT[J]. J Magn Reson Imaging, 2008, 28(2): 440-444.
[12]李鹏,吕发金,勒都晓兰,等. 匀场辅助装置在臂丛神经成像中的应用[J]. 中国医学影像技术,2012,28(10):1906-1909.

Accesses

Citation

Detail

段落导航
相关文章

/